GE Venue 50 Basic Service Manual page 261

Hide thumbs Also See for Venue 50:
Table of Contents

Advertisement

Peripheral/option checks
Mains cable inspection
Table 10-7:
Step
Item
1.
Unplug Cord
2.
Inspect
3.
Verify
Cleaning
Step
Item
1.
Console
2.
Probe Holder
3.
LCD
Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
If any peripherals or options are not part of the system
configuration, the check can be omitted.
See 'Functional checks' on page 4-8 for more information.
Mains Cable Inspection, As Appropriate
Disconnect the mains cable from the wall and Ultrasound system.
Inspect it and its connectors for damage of any kinds.
Verify that the LINE, NEUTRAL and GROUND wires are properly attached
to the terminals, and that no strands may cause a short circuit.
Table 10-8:
General Cleaning
Remove the battery. Use a fluid detergent in warm water on a soft, damp
cloth to carefully wipe the entire system. Be careful not to get the cloth too
wet so that moisture does not enter the console.
Clean probe holders. (they may need to be soaked to remove excess gel).
Use a soft, non-abrasive folder cloth or use the clearners listed in Chapter
6 of User Manual. Gently wipe the LCD face. DO NOT use a glass cleaner
that has a hydrocarbon base (such as Benzene, Methy Alcohol or Methy
Ethyl Ketone) on LCD with the filter (anti-glare shield).
System maintenance
Description
Description
10-13