Electrical Requirements - GE Venue 50 Basic Service Manual

Hide thumbs Also See for Venue 50:
Table of Contents

Advertisement

Site Preparations

Electrical requirements

General requirements
NOTE:
NOTE:
Electrical requirements for the Ultrasound system
Table 2-3:
Voltage
100-240 VAC
Table 2-4:
Voltage
100-240 VAC
Inrush Current
2-6
GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full size ground wire from
the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full size ground wire from the
distribution panel to the Ultrasound outlet.
Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound system is only a conduit.
Electrical Specifications for Venue 50 system
Power
180 VA max.
Electrical Specifications for Docking Cart
Power
380 VA max.
Inrush current is not a factor to consider due to the inrush
current limiting properties of the power supplies.
Table 2-5:
Voltage
100V
240V
Current
1.6A (max)
Current
2.7A (max)
Inrush Current
Inrush Current
75A (max)
120A (max)
Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
Frequency
50/60HZ
Frequency
50/60HZ