Electrical Safety - GE Venue 50 Basic Service Manual

Hide thumbs Also See for Venue 50:
Table of Contents

Advertisement

Introduction

Electrical safety

Safe practices
WARNING
Probes
Peripherals
1-20
Follow these guidelines to minimize shock hazards whenever
you are using the Ultrasound system:
To minimize shock hazard, the equipment chassis must be
connected to an electrical ground.
The Ultrasound system is equipped with a three-conductor
AC power cable. This must be plugged into an approved
electrical outlet with safety ground.
The power outlet used for this equipment should not be
shared with other types of equipment.
Both the system power cable and the power connector must
meet international electrical standards
Connecting a Venue 50 to the wrong voltage level will most
likely destroy it.
Follow these guidelines before connecting a probe to the
Ultrasound system:
Inspect the probe prior to each use for damage or
degradation to the:
housing
cable strain relief
lens
seal
connector pins
locking mechanism
Do not use a damaged or defective probe.
Never immerse the probe connector or adapter into any
liquid.
Refer to the Patient Safety Environment section of the User's
Manual for peripheral isolation information.
Venue 50 – Basic Service Manual
5447566-100 English Rev. 7