GE Venue 50 Basic Service Manual page 148

Hide thumbs Also See for Venue 50:
Table of Contents

Advertisement

General Procedures and Functional Checks
Basic Measurements
NOTE:
Distance Measurements
Circumference/Area (Ellipse) Measurement
M Mode Measurements
4-20
The following instructions assume that you first scan the patient
and then press Freeze.
1. Press Measure.
2. Select Distance.
3. Touch the screen, and two active calipers display.
4. Move the calipers to the desired position.
5. Press Set, then the value is displayed in the results window.
1. Press Measure.
2. Select Ellipse.
3. Touch the screen, and three active calipers display.
4. Move the calipers to the desired position.
5. Press Set, then the value is displayed in the results window.
1. Press Measure.
2. Select Depth or Heart Rate.
3. Touch the screen and two active calipers display.
4. Move the calipers to the desired position.
5. Press Set, then the value is displayed in the results window.
Venue 50 – Basic Service Manual
5447566-100 English Rev. 7