When There's Too Much Leakage Current; Ac/Dc Fails; Enclosure Fails - GE Venue 50 Basic Service Manual

Hide thumbs Also See for Venue 50:
Table of Contents

Advertisement

Care and Maintenance
When there's too much leakage

AC/DC Fails

Enclosure Fails

NOTE:
10-32
Where applicable, check the AC/DC adapter and its cable.
Replace a new one if any portion is defective.
Check for any damage to the enclosure. Replace any defective
part.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other
wiring abnormalities. Notify the user or owner to correct any
deviations. As a work around, check the other outlets to see if
they could be used instead.
No outlet tester can detect the condition where the white neutral
wire and the green grounding wire are reversed. If later tests
indicate high leakage currents, this should be suspected as a
possible cause and the outlet wiring should be visually
inspected.
current ...