Site circuit breaker
CAUTION
Site power outlets
Unit power plug
Power stability requirements
Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
Power outage may occur. The Venue 50 requires a dedicated
single branch circuit. To avoid circuit overload and possible loss
of critical care equipment, make sure you do not have any
other equipment operating on the same circuit.
It is recommended that the branch circuit breaker for the
Ultrasound system be readily accessible.
A dedicated AC power outlet must be within reach of the
Ultrasound system without extension cords. Other outlets
adequate for the external peripherals, medical and test
equipment needed to support this Ultrasound system must also
be present within 1 m (3.2 ft.) of the Ultrasound system.
Electrical installation must meet all current local, state, and
national electrical codes.
If the Ultrasound system arrives without a power plug, or with
the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
Voltage drop-out:
Max 10 ms.
Power transients (all applications):
Less than 25% of nominal peak voltage for less than 1
millisecond for any type of transient, including line frequency,
synchronous, asynchronous, or aperiodic transients.
General Requirements
2-7