4700 Refrigerated Sampler Service Guide
Section 3 Refrigeration System
Figure 3-3 Control panel module removal
Figure 3-4 Temperature sensor cable wiring connector
3-4
3. Lift the control panel to expose the wiring connectors, tak-
ing care not to put strain on the wires.
4. Disconnect the temperature sensor cable from connector
P5 on the control board (Figure 3-4) and set the panel back
in place.
When disconnecting a cable connector, always grasp the con-
nector itself and not the wires.
P5
5. Slide the unmarked sensor out of its black plastic holder
(located behind the power supply plate, Figure 3-1,
Detail B).
6. Remove the marked sensor from the evaporator coil and
rear wall by cutting the two cable ties holding it in place.
7. The wiring assembly enters the rear of the refrigeration
cabinet just above the upper left corner of the
CAUTION