2.1 The radEAGLET-R Display
Battery Status
The battery status is shown by a value and a symbol.
Full battery (100 %)
Empty battery
While the radEAGLET-R is powered by an external source (see 8.3, p. 123)
providing enough power to charge the internal battery, the symbol is deco-
rated by a bolt.
 
While the radEAGLET-R is powered by an external source (see 8.3, p. 123)
and there is no or a non-rechargeable battery installed, this symbol is
shown.
Connectivity Status
This item is available while a communication dongle is plugged into the
radEAGLET-R's USB-A port.
The radEAGLET-R is logged into a Wi-Fi network.
The radEAGLET-R provides a Wi-Fi hotspot other devices can connect to.
The radEAGLET-R is logged into a wired local area network ("Ethernet").
The radEAGLET-R communicates via Bluetooth.
Communication is off.
A USB mass storage device is available.
GPS Status
This item is available for radEAGLET-R models equipped with a GPS receiver
(see E.1, p. 153).
GPS is switched on and receives enough data from several satellites to
calculate the position ("has fix").
Location data is included whenever you save measuring results, for exam-
ple, a spectrum.
GPS is switched on but reception is (hitherto) insufficient.
GPS is switched off.
Stabilization Status
The stabilization (see C.1, p. 147) status is indicated by this symbol.
(Flashing) Stabilization is in progress. After about 2–3 minutes, the status
should change to one of the following.
The instrument is stabilized. You can use the instrument.
The continous stabilization process got disturbed. See appendix D.1,
p.
151
Data Logging
This symbol shows whether the radEAGLET-R logs data.
24/163
     
     
for recommended remedies.
Software 3.2.12r • Document 3.3.3b4o
 
 
radEAGLET-R User Manual