radEAGLET-R User Manual
9.1.2 Calibrate the radEAGLET-R
Usually the calibration and stabilization of the radEAGLET-R happens during startup and
continuously while using it. If you want to explicitly start a calibration with a reference nuclide,
several methods are available. A more elaborate method is described in chapter 5.18, p. 76.
Here we use an automated calibration procedure.
For additional information refer to appendices C.1, p.
(calibration), and D.1, p.
To calibrate the radEAGLET-R...
1. Move to a low-radiation environment.
2. If your radEAGLET-R has no calibration source built-in, place a source in front of the
detector, for example, the KCl box included with the instrument.
3. Open the
Advanced
4. Scroll down to
5. Specify to the calibration nuclide and
6. The instrument displays progress in percent (%) during the process (Fig. 137, p. 130).
7. Wait until the radEAGLET-R displays "System is operative" (Fig. 138, p. 130).
8. Finish the calibration with
© innoRIID GmbH • 2020-11-18
General Information . . . . . . . . . . . . . . . see 5.28, p.
Available Memory . . . . . . . . . . . . . . . . . see 5.22, p.
GPS
. . . . . . . . . . . . . . . . . . . . . . . . . . see 5.13, p.
Self Test . . . . . . . . . . . . . . . . . . . . . . . . . see 5.27, p.
Hardware Status . . . . . . . . . . . . . . . . . see 5.16, p.
Figure 136:

Commands for status information

151
(red stabilization icon).
menu.
Easy Calibration
Exit .
Software 3.2.12r • Document 3.3.3b4o
147
and
Select
it.
Select
it.
92
82
70
90
74
(stabilization), C.2, p.
129/163