Ametek 270SF Instruction Manual
Ametek 270SF Instruction Manual

Ametek 270SF Instruction Manual

Extractive gas sample probe
Hide thumbs Also See for 270SF:

Advertisement

Instruction Manual
Model 270SF
Extractive Gas Sample Probe
5200 Convair Drive Carson City, NV 89706 • Phone: 775-883-2500 • Fax: 775-883-6388 • www.universalanalyzers.com
MAN.270SF.REVA.12222014

Advertisement

Table of Contents
loading

Summary of Contents for Ametek 270SF

  • Page 1 Instruction Manual Model 270SF Extractive Gas Sample Probe 5200 Convair Drive Carson City, NV 89706 • Phone: 775-883-2500 • Fax: 775-883-6388 • www.universalanalyzers.com MAN.270SF.REVA.12222014...
  • Page 2: Table Of Contents

    Contents Receiving and Storage Definition of Symbols Product Identification Specifications Description and Principle of Operation Installation Electrical Connections Process and Piping Connections Start-Up Shutdown Maintenance Troubleshooting Spare Parts Drawings - Model 270SF Limited Warranty Page 2 of 45 MAN.270SF.REVA.12222014...
  • Page 3: Receiving And Storage

    Receiving and Storage The UAI 270SF Extractive Gas Sample Probe is a complete pre-installed unit. No assembly is necessary when received on-site. Carefully inspect the product and any special accessories included with it immediately on arrival by removing them from the packing and checking for missing articles against the packing list.
  • Page 4: Definition Of Symbols

    SHOULD BE TAKEN BEFORE DISCONNECTING OR CONTACTING ANY ELECTRICAL CONNECTIONS. PROTECTIVE CONDUCTOR TERMINAL SYMBOL INDICATES THE TERMINAL LOCATION FOR THE PROTECTIVE CONDUCTOR. FAILURE TO CONNECT TO THE PROTECTIVE CONDUCTOR TERMINAL MAY RESULT IN A SHOCK HAZARD. Page 4 of 45 MAN.270SF.REVA.12222014...
  • Page 5: Product Identification

    2 µm Sintered 316SS Alarm Output Closed on Alarm Fail Safe (Standard) TK Type K ThermoCouple RD 100 Ohm RTD None -TS340 -SBB -115 -115 -CM Sample Part # NOTE: Probe Tip Filters & Probe Tubes Purchased Separately. Page 5 of 45 MAN.270SF.REVA.12222014...
  • Page 6: Specifications

    4” Heat shrink boot (optional) 2” Heat shrink boot (optional) Heater Type Rod heaters in aluminum block, controlled w/ thermal switch Available Filter 2 µm ceramic (standard) 316SS (optional) Chamber Materials 316SS (standard) 316SS, SilcoNert™ coated Hastelloy C-276 Page 6 of 45 MAN.270SF.REVA.12222014...
  • Page 7: Description And Principle Of Operation

    The 270SF filter will mount by means of a pipe flange to a mating flange on the stack. The size of the flange can be specified by the user and can include 2”, 3”, 4”, 6” or DIN type. A 1/2” NPT female threaded connection is provided for the sample probe to mount inside the mounting flange.
  • Page 8: Installation

    After securing the heated sample line, use a heat gun to shrink the entry boot onto the sample line. The goal is to keep the temperature of the gas sample at a temperature above the dew point of the gas all the way from the stack to the sample cooler/dehydrator. Page 8 of 45 MAN.270SF.REVA.12222014...
  • Page 9: Electrical Connections

    Electrical Connections Model 270SF Non-Heated Enclosure Configuration w/o Blowback Page 9 of 45 MAN.270SF.REVA.12222014...
  • Page 10 Electrical Connections Model 270SF back Heated Enclosure Configuration w/o Blow Heated Enclosure w/ Blowback Page 10 of 45 MAN.270SF.REVA.12222014...
  • Page 11 Electrical Connections Model 270SF 230V 230V 115V 115V 115V 115V 230V 230V 115V 115V 115V 115V 230V 115V 115V Page 11 of 45 MAN.270SF.REVA.12222014...
  • Page 12 Electrical Connections Model 270SF 230V 230V 115V 115V 115V 115V Page 12 of 45 MAN.270SF.REVA.12222014...
  • Page 13 Electrical Connections Model 270SF 230V 230V 230V 230V 115V 115V 115V 115V 115V 115V 115V 115V Page 13 of 45 MAN.270SF.REVA.12222014...
  • Page 14 Electrical Connections Model 270SF 230V 230V 230V 230V 115V 115V 115V 115V 115V 115V 115V 115V Page 14 of 45 MAN.270SF.REVA.12222014...
  • Page 15 Electrical Connections Model 270SF 230V 230V 230V 230V 115V 115V 115V 115V 115V 115V 115V 115V Page 15 of 45 MAN.270SF.REVA.12222014...
  • Page 16: Process And Piping Connections

    Process and Piping Connections Model 270SF Page 16 of 45 MAN.270SF.REVA.12222014...
  • Page 17: Start-Up

    6. After purging is complete, follow the maintenance procedure to change the filter. 7. Cap the sample outlet tube connection and disconnect power from the unit. Note: If electrical wires are to be disconnected, follow applicable ‘Lock Out/ Tag Out’ requirements. Page 17 of 45 MAN.270SF.REVA.12222014...
  • Page 18: Maintenance

    Replace the filter with a new one, again handling it with pliers. Ensure that the filter is pushed in the center of the oven so that it is in contact with the O-Ring on the far end of the filter. Screw the cap back on the filter body. The filter replacement procedure is complete. Page 18 of 45 MAN.270SF.REVA.12222014...
  • Page 19: Troubleshooting

    Leaking blowback solenoid valve. Block or disconnect the blowback supply Loose connection Verify all fittings are leak free Low readings during calibration Insufficient calibration gas flow Ensure calibration flow is at least 110% of the sample gas flow Page 19 of 45 MAN.270SF.REVA.12222014...
  • Page 20: Spare Parts

    Solenoid Valve Blowback, 2-Way 240VAC 4955-0005 Check Valve Calib. Gas, Adjustable Pressure 5150-0014 Accumulator Sphere Blowback, Unheated 4956-0001 Acid Blocker Assembly, Blowback 5110-0015 Gasket Ring For 4” 150# Flange Garlock 9850 1/16” Thk 4903-0001 * Commissioning Spare Part Page 20 of 45 MAN.270SF.REVA.12222014...
  • Page 21: Drawings - Model 270Sf

    Drawings Model 270SF Page 21 of 45 MAN.270SF.REVA.12222014...
  • Page 22 Drawings Model 270SF Page 22 of 45 MAN.270SF.REVA.12222014...
  • Page 23 Drawings Model 270SF Page 23 of 45 MAN.270SF.REVA.12222014...
  • Page 24 Drawings Model 270SF Page 24 of 45 MAN.270SF.REVA.12222014...
  • Page 25 Drawings Model 270SF Page 25 of 45 MAN.270SF.REVA.12222014...
  • Page 26 Drawings Model 270SF Page 26 of 45 MAN.270SF.REVA.12222014...
  • Page 27 Drawings Model 270SF Page 27 of 45 MAN.270SF.REVA.12222014...
  • Page 28 Drawings Model 270SF Page 28 of 45 MAN.270SF.REVA.12222014...
  • Page 29 Drawings Model 270SF Page 29 of 45 MAN.270SF.REVA.12222014...
  • Page 30 Drawings Model 270SF Page 30 of 45 MAN.270SF.REVA.12222014...
  • Page 31 Drawings Model 270SF Page 31 of 45 MAN.270SF.REVA.12222014...
  • Page 32 Drawings Model 270SF Page 32 of 45 MAN.270SF.REVA.12222014...
  • Page 33 Drawings Model 270SF Page 33 of 45 MAN.270SF.REVA.12222014...
  • Page 34 Drawings Model 270SF Page 34 of 45 MAN.270SF.REVA.12222014...
  • Page 35 Drawings Model 270SF Page 35 of 45 MAN.270SF.REVA.12222014...
  • Page 36 Drawings Model 270SF Page 36 of 45 MAN.270SF.REVA.12222014...
  • Page 37 Drawings Model 270SF Page 37 of 45 MAN.270SF.REVA.12222014...
  • Page 38 Drawings Model 270SF Page 38 of 45 MAN.270SF.REVA.12222014...
  • Page 39 Drawings Model 270SF Page 39 of 45 MAN.270SF.REVA.12222014...
  • Page 40 Drawings Model 270SF Page 40 of 45 MAN.270SF.REVA.12222014...
  • Page 41 Drawings Model 270SF Page 41 of 45 MAN.270SF.REVA.12222014...
  • Page 42 Drawings Model 270SF Page 42 of 45 MAN.270SF.REVA.12222014...
  • Page 43 Drawings Model 270SF Page 43 of 45 MAN.270SF.REVA.12222014...
  • Page 44: Limited Warranty

    UAI, UAI will only be bound to comply with the length of time associated with such warranty. All other terms of such warranty will be governed by this Standard Limited Warranty and UAI’s General Terms and Conditions incorporated herein by reference. Page 44 of 45 MAN.270SF.REVA.12222014...
  • Page 45 (ii) notified UAI of the limited warranty claim within sixty (60) days from the date Buyer knew or had reason to know of the defect 5200 Convair Drive Carson City, NV 89706 • Phone: 775-883-2500 • Fax: 775-883-6388 • www.universalanalyzers.com Page 45 of 45 MAN.270SF.REVA.12222014...

Table of Contents