Contents Receiving and Storage Definition of Symbols Product Identification Specifications Description and Principle of Operation Installation Electrical Connections Process and Piping Connections Start-Up Shutdown Maintenance Troubleshooting Spare Parts Drawings - Model 270SF Limited Warranty Page 2 of 45 MAN.270SF.REVA.12222014...
Receiving and Storage The UAI 270SF Extractive Gas Sample Probe is a complete pre-installed unit. No assembly is necessary when received on-site. Carefully inspect the product and any special accessories included with it immediately on arrival by removing them from the packing and checking for missing articles against the packing list.
SHOULD BE TAKEN BEFORE DISCONNECTING OR CONTACTING ANY ELECTRICAL CONNECTIONS. PROTECTIVE CONDUCTOR TERMINAL SYMBOL INDICATES THE TERMINAL LOCATION FOR THE PROTECTIVE CONDUCTOR. FAILURE TO CONNECT TO THE PROTECTIVE CONDUCTOR TERMINAL MAY RESULT IN A SHOCK HAZARD. Page 4 of 45 MAN.270SF.REVA.12222014...
The 270SF filter will mount by means of a pipe flange to a mating flange on the stack. The size of the flange can be specified by the user and can include 2”, 3”, 4”, 6” or DIN type. A 1/2” NPT female threaded connection is provided for the sample probe to mount inside the mounting flange.
After securing the heated sample line, use a heat gun to shrink the entry boot onto the sample line. The goal is to keep the temperature of the gas sample at a temperature above the dew point of the gas all the way from the stack to the sample cooler/dehydrator. Page 8 of 45 MAN.270SF.REVA.12222014...
6. After purging is complete, follow the maintenance procedure to change the filter. 7. Cap the sample outlet tube connection and disconnect power from the unit. Note: If electrical wires are to be disconnected, follow applicable ‘Lock Out/ Tag Out’ requirements. Page 17 of 45 MAN.270SF.REVA.12222014...
Replace the filter with a new one, again handling it with pliers. Ensure that the filter is pushed in the center of the oven so that it is in contact with the O-Ring on the far end of the filter. Screw the cap back on the filter body. The filter replacement procedure is complete. Page 18 of 45 MAN.270SF.REVA.12222014...
Leaking blowback solenoid valve. Block or disconnect the blowback supply Loose connection Verify all fittings are leak free Low readings during calibration Insufficient calibration gas flow Ensure calibration flow is at least 110% of the sample gas flow Page 19 of 45 MAN.270SF.REVA.12222014...