9.2 Survey the Environment
2. The radEAGLET-R always compares the current dose rate with the thresholds speci-
fied for warnings and alarms (see 5.17, p. 75) and reports values above the thresholds
(Fig. 141, p. 132).
Press
Confirm
details.
9.2.2 Neutron Radiation
This item is available for radEAGLET-R models with a neutron detector (see E.1,
p. 153).
The radEAGLET-R's neutron detector can detect slow a.k.a. thermal neutrons.
Fast neutrons need to be slowed down by a moderator, for example, ordinary water or sub-
stances containing lots of hydrogene.
To survey for neutron radiation...
1. Increase the sensitivity for unmoderated neutrons by holding the radEAGLET-R so the
neutron detector is close to your body (Fig. 142, p. 133).
2. To detect neutrons regardless of your position relative to the source turn yourself
around.
3. Observe the screen. Neutron incidents are shown and counted (Fig. 143, p. 133).
4. If the neutron counts exceed the adjustable alarm threshold (see 5.17, p. 75) a neu-
tron alarm is triggered (see 4.1.1, p. 45).
For unmanned neutron detection a moderator must be used additionally. See Ap-
pendix E.4, p.