Figure 4–8. Count Results, Soil Mode
✓ The PDA stores the measurement results in the active project,
then displays the confirmation message Soil results stored
in project: XXX, where XXX is the active project name. Press
the 〈ok〉 button to continue.
✓ The PDA then stores the project data in the gauge. Upon
completion, the PDA displays Project storage to gauge
complete. Press the 〈ok〉 button to return to the main menu.
If the gauge is unable to obtain GPS data after taking a
count, it will still send the count results to the PDA and
the results will be stored in the active project.
✓ Lift the gauge from the test site by the source rod handle. This
returns the source rod to the SAFE (shielded) position. When
not taking readings, always keep the source rod in the SAFE
position. If you do not hear a click when the gauge is raised to
the SAFE position, look at the bottom of the gauge to verify
that the tungsten sliding block is completely closed. If the gauge
base opening is not completely closed by the sliding block, the
sliding block may require cleaning. Refer to page A–11 for
cleaning instructions.