✓ Ensure the source rod handle is fully latched in the SAFE
(shielded) position.
✓ Check that the gauge is positioned correctly on the reference
standard block (see Figure 4–2 on page 4–3).
✓ Perform the standard count again. If the standard count still
fails, contact the nearest Troxler Service Center.
GAUGE READINGS APPEAR ERRATIC
✓ If taking readings in the backscatter position, ensure the source
rod is properly locked in the notch and not resting on the test
material.
✓ Check the inside of the gauge for moisture. If necessary, use a
hair dryer (on low heat) to circulate warm air for one to three
hours.
✓ Remove any foreign objects from inside the gauge.
✓ Ensure the hardware mounting screws are tight and in place.
✓ Check count time – a four-minute count will give the highest
precision, with a repeatability of ±1 pcf on repeated tests
performed without moving the gauge.
✓ Erratic density readings may be caused by a dirty tungsten
sliding block. Clean the sliding block (see page A–11).
✓ Perform a stat test:
If test passes, proceed with job.
If test fails, repeat two more times. If test fails two out of
three times, contact the nearest Troxler Service Center.
GAUGE TURNS OFF AFTER IT IS TURNED ON
✓ If the rechargeable battery is below 5.5 volts, recharge or
replace the batteries (see page C–24).
Model 3451
Appendix C–11
Need help?
Do you have a question about the Enhanced RoadReader Plus 3451 and is the answer not in the manual?