Troxler Enhanced RoadReader Plus 3451 Manual Of Operation And Instruction page 198

Surface moisture-density gauge
Table of Contents

Advertisement

Error! Gauge communication will NOT be possible!
This error can occur if the gauge and the PDA cannot communicate.
Possible causes:
✓ Gauge not turned on.
✓ PDA is not within 4.6 m (15 ft) of the gauge.
✓ Object in the way of the transmitter/receiver.
✓ Check that the PDA is the one shipped with the gauge. Before
leaving the Troxler factory, the transceivers on the gauge and
PDA are programmed with a unique, matching address. The
PDA can only communicate with the matching gauge.
✓ Verify the PDA's Bluetooth communications configuration as
described on page 3–14.
✓ Verify the PDA's ActiveSync configuration as described on
page 3–13.
ERROR! Invalid ABORT Characters.
This error can occur when trying to abort from a process.
Possible causes:
✓ PDA is not within 4.6 m (15 ft) of the gauge.
✓ Object in the way of the transmitter/receiver.
Error! Invalid data received.
This error can occur if a command or data transmitted to the gauge
or received by the PDA is not legal.
Possible causes:
✓ PDA is not within 4.6 m (15 ft) of the gauge.
✓ Object in the way of the transmitter/receiver.
ERROR! Invalid Depth Characters.
This error can occur when trying to recalibrate the depth strip from a
process.
Possible causes:
✓ PDA is not within 4.6 m (15 ft) of the gauge.
✓ Object in the way of the transmitter/receiver.
Model 3451