STAT TEST

As described on page C–2, erratic readings, or readings that seem to
fluctuate, may indicate a problem with the gauge. If the readings are
suspect, the operator can perform a stat test, or statistical stability
test, to validate the normal operation of the gauge.
To view the results of the most recent stat test, press the
〈STAT Test〉 button on the Print Active Project Menu. As
shown in Figure 7–12, the Data Manager software displays the stat
test results.

DRIFT TEST

As described on page C–5, if the gauge passes the stat test, but
shows long-term drift between measurements, the operator can
perform a drift test to check the long-term drift.
To view the results of the most recent drift test, press the
〈Drift Test〉 button on the Print Active Project Menu. As shown
in Figure 7–13, the Data Manager software displays the drift test
results.

SEND TO PRINTER

To print the configuration of the active project, press the
〈Send to Printer〉 button on the Print Active Project Menu.
The software prompts the operator to ensure that a printer is
connected to the computer and that the printer is powered on. After
doing so, press 〈OK〉. The software sends the project configuration
to the default printer.

SEND TO FILE

To save the configuration of the active project as a comma-
delimited text (.txt) file, press the 〈Send to File〉 button on the
Print Active Project Menu. In the resulting display, enter a file
name and press 〈Open〉 to save the file.