3.7.9 Alarms check

Low Pressure
High Pressure
3-44
WARNING !! Severe patient injuries
The alarms must trigger at the proper time and in the correct manner.
Check the proper activation of the visual and acoustic signals.
Before starting the alarm checking:
check the alarms setup, and change the set values whenever
necessary (please see previous paragraph)
select VC/VAC operative mode
Set the low pressure alarm limit to 5 cmH 2 O.
Disconnect the patient simulator from the patient circuit during
ventilation.
After about 20 seconds the system activates the airways low
pressure alarm: silence the alarm.
Reconnect the patient simulator.
Set the high pressure alarm limit to a value higher than the PAW
ventilation pressure by 5 cmH 2 O.
Block the patient simulator (using your hands) during ventilation.
The system activates the airways high pressure alarm: silence the
alarm.
Unlock the patient simulator.
Press the START button: the
lung ventilator begins its cycle.
User manual, DU30651043