3.7.9 Alarms check
Low Pressure
High Pressure
3-44
WARNING !! Severe patient injuries
•
The alarms must trigger at the proper time and in the correct manner.
•
Check the proper activation of the visual and acoustic signals.
Before starting the alarm checking:
•
check the alarms setup, and change the set values whenever
necessary (please see previous paragraph)
•
select VC/VAC operative mode
•
Set the low pressure alarm limit to 5 cmH 2 O.
•
Disconnect the patient simulator from the patient circuit during
ventilation.
•
After about 20 seconds the system activates the airways low
pressure alarm: silence the alarm.
•
Reconnect the patient simulator.
•
Set the high pressure alarm limit to a value higher than the PAW
ventilation pressure by 5 cmH 2 O.
•
Block the patient simulator (using your hands) during ventilation.
•
The system activates the airways high pressure alarm: silence the
alarm.
•
Unlock the patient simulator.
•
Press the START button: the
lung ventilator begins its cycle.
User manual, DU30651043