5.4

Troubleshooting

This chapter is an indicative but not exhaustive guide for the user and the technician,
providing indications for eliminating, as quickly as possible, most of the problems that may
have caused malfunctioning or alarm signals.
This chapter describes the possible causes of problems, indicated by alarms that are
activated during normal functioning.

5.4.1 Troubleshooting list

Switch ON failure
Power Failure
Siaretron4000e
WARNING !!
If the problem persists, carry out a complete check of the lung ventilator to
identify any irregularities.
If the problem cannot be resolved, contact the Siare Service Centre or a
Centre authorized by Siare.
The lung ventilator does not switch on
Check that it is connected to the main power supply.
Check that the main switch is turned to the I position (ON).
Check the main fuses.
Check the display connection cable.
Contact the Siare Service Centre or a Centre authorized by Siare.
There is a power supply fault and the lung ventilator is operating on
the battery
Check if the LED power indicator is ON.
Check that it is connected to the main power supply.
Check that the main switch is turned to the I position (ON).
Check the correct connections of the plug, the fuses and the
connector, and the condition of the cable (if necessary, restore
the connections and replace the cable if it is damaged).
Check that power is present at the relative socket by plugging in
another electrical device. If there is no power, use another socket
or check the overload switch on the electrical panel of the room.
Check the 12V voltage is properly supplied to the main board.