4.8.11 APRV (Airway Pressure Release Ventilation)

Airway ventilation and pressure release: this type of
ventilation features two positive pressure levels.
The system displays all PRP relative to the set
operating mode.
APRV is a spontaneous ventilation mode with constant positive pressure on 2 different
levels, during which the lung ventilator keeps the two pressure levels set constant.
In this operating mode the patient is free to breath spontaneously inside the circuit but at a
pressure greater than the atmospheric one, with increased residual functional capacity.
The clinician can set on the lung ventilator the pressure of the two levels and the relative
times.
LOW LEVEL pressure - LOW level TIME
HIGH LEVEL pressure - HIGH level TIME
4-34