3.7.3 Leak Test
TESTS ON DEMAND. Leak Test function is
selected.
Siaretron4000e
WARNING !! Risk of failure of ventilator and/or injury to the patient.
This test verifies that there are no leaks higher than 100 ml/min in the lung
ventilator pneumatic circuits.
WARNING !! Risk of injury for the patient.
Perform the Leak Test weekly and/or when patient circuit is changed.
•
Press the knob; the system will
activate the leak test.
•
Close the patient circuit (close
manually the patient connector).
•
Press START to start the
procedure.
•
The Leak Test procedure is in
progress.
3-35