3.9.1 Respiratory flow sensors calibration

The respiratory flow sensors calibration is necessary in the following cases:
noted differences are more than 15% ( over 100ml ) between the set Volume value
( VTi - Vte ) and the expired Tidal Volume reading ( Vte ).
in case of first calibration of ventilator or replacement of flow sensors (INSP inside
the unit or EXP outside the unit), it is suggested to perform this Calibration.
3-50
In order to start the respiratory flow sensors calibration program is
necessary the intervention of qualified SIARE personnel or qualified
technical personnel authorized by SIARE.
Turn the encoder knob to
select the parameter to be
calibrated: the activated
message is highlighted.
Press the encoder knob: the
respiratory flow sensors
calibration program is
showed.