3.3.5 Protection fuses
Protection fuses are foreseen in the following
circuit:
•
Main power supply (protection fuses
without compressor: 2x3,15 AT).
•
Lung ventilator power supply fuses
(2x1 AT).
•
External 12Vdc power supply fuse
(6,35 AT).
WARNING!! Risk of injury for the user / patient
The operations described here below must be carried out by highly qualified
technical personnel, trained and formally authorised by SIARE only.
In case of break of protection fuse:
•
interrupt the main power supply;
•
eliminate the malfunctioning or the cause which caused the fuse break,
•
replace the protection fuse with one with the same value and technical characteristics.
WARNING!! Risk of injury for the user / patient
Fuses with wrong values and technical characteristics can compromise
equipment integrity and lung ventilator safety.
3-14
User manual, DU30651043