A.5 Electromagnetic compatibility tables
A.5.1 Annex A - Table 1
Guidance and manufacturer's declaration – electromagnetic emissions
The Siaretron 4000 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Siaretron 4000 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
A-18
Compliance
Virdict
Group 1
Compliance
Class A
Compliance
Class A, B, C, D, o
Class A
NOT APPLICABLE
Compliance
Compliance
Electromagnetic environment – guidance
The Siaretron 4000 uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The Siaretron 4000 is suitable for use in
all establishments other than domestic,
and may be used in domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes,
provided the following warning is heeded:
Warning:
This equipment/system is
intended for use by healthcare
professionals only. This equipment/
system may cause radio interference or
may disrupt the operation of nearby
equipment. It may be necessary to take
mitigation measures, such as re-orienting
or relocating the Siaretron 4000 or
shielding the location.
User manual, DU3065104