3.3.8 Use of antibacterial filter

Apply the antibacterial filters to the patient circuit.
3-16
WARNING!! Risk of injury for the patient.
To protect the patient from particles and dust, it is necessary to use a filter
between the inspiratory hose and the patient, i.e., the filter on the Y
connector or the filter on inspiratory hose.
WARNING!! Risk of injury for the patient.
Replace the antibacterial filters as indicated in the maintenance
instructions.
WARNING!! Risk of injury for the patient.
Please perform the "SELF TEST" phase (lung ventilator - switch ON) every
time you replace the patient circuit.
The system will check the patient circuit every time you switch ON the lung
ventilator.
WARNING!! Risk of strangling.
Pay special warning when connecting the patient circuit to the lung
ventilator.
If not carefully positioned, the hoses, the cables, the patient circuit and
other similar components to the lung ventilator, these can be dangerous for
the patient.
WARNING!! Risk of burns
Do not use conductive masks or conductive breathing hoses during surgery
with electrosurgical units because they can cause burns.
User manual, DU30651043