3.3

Preparation to use

3.3.1 Medical gas connection

Screw the AIR gas supply hose to relevant
connectors of the gas supply area on the
trolley.
Screw the O 2 gas supply hose to relevant
connectors of the gas supply area on the
lung ventilator.
Screw the gas supply hoses to the relevant connectors of the hospital distribution
system. Be sure that the pneumatic hoses are well fixed and that the hospital medical
gas pipeline system is clean and without lubricants.
3-8
The pneumatic hoses are already supplied with screw connection DISS
type (Diameter Index Safety System) for connection to the lung ventilator
connectors.
The installation technician taking care of the hoses connection to the
outlets must ensure that they are compatible with the hospital medical gas
pipeline system.
To prevent inversions of gases that can be FATAL for the patient, assembly
of the connectors compatible with the hospital distribution system and all
the maintenance and/or replacement operations of the medical gas supply
hoses must be carried out by highly qualified technical personnel only.
The lung ventilator can also works with AIR medical gas supply only, but in
this case the FiO 2 will be adjustable at 21% only. An alarm indicating "Low
Oxygen Pressure" condition will be activated.