A2 - ALARMS parameter, MAIN MENU

This monitor area, through the encoder knob ( MAIN MENU - ALARMS ), allows to display
and set the Min and Max. alarms values.
Low Pressure
High Pressure
Low Resp. Rate
High Resp. Rate
Low Exp. Vte
High Exp. Vte
Low Exp. VM
High Exp. VM
Low PEEP
High PEEP
Low FiO 2
High FiO 2
Gas Sensor : Low EtCO 2
Gas Sensor : High EtCO 2
Power Failure
5-6
WARNING!! Risk of injury for the patient
The lung ventilator used in the same health environments can have
different preset configurations of alarm limits.
Verify that the preset alarm limits are appropriate for the new patient
and adjust the alarm limits on values suitable to the new condition of
use.
The alarms setting to the range limit could make the alarm not properly
working.
From 2 to 60 cmH
O
2
From 20 to 81 cmH
O
2
From 1 to 11 bpm
From 13 to 150 bpm
From 0 to 1400 ml (10 ml steps)
From 1 to 3000 ml (10 ml steps)
From 0 to 20 L
From 1 to 100 L
From 0 to 19 cmH
O
2
From 1 to 51 cmH
O
2
From 20 to 98 %
From 21 to 100 %
From 0.4 to 12.9 %
From 0.5 to 13.0 %
Enable / Disable
User manual, DU30651043