5.2.4 A4 - Acoustic alarm silencing
•
•
The soft key, during the normal operating phase of lung
ventilator, allows the silencing of active acoustic alarm.
WARNING!! Risk of injury for the user / patient
The user should never stop checking the patient conditions during the alarm
silencing.
•
Pushing the RESET button will stop the acoustic alarm
for a defined time.
•
During the alarm silencing the alarm text is showed on
the panel.
•
Pushing further the RESET button will cancel the alarm
text only if the alarm conditions is disappeared.
•
If during the alarm silencing a new high priority alarm
occur, the alarm silencing is cancelled and the acoustic
signal and the visual texts are activated again.
5-8
It's possible to change the alarm settings even when the alarms are
activated.
After changing an alarm setting, the relevant sign is lighted and the
status icon will flash for a defined time.
User manual, DU30651043