3.3.13 Connection to other equipments

Siaretron4000e
Connection to Siare equipment
If the equipment to be connected is a SIARE unit, all the instructions
necessary for the connection to lung ventilator can be found in the
documentation supplied.
WARNING!! Risk of injury for the user / patient
Do not connect external devices NOT manufactured or NOT authorized by
SIARE to the lung ventilator (e.g., scavenging systems, patient simulators),
and not described in the present user's manual.
In case contact SIARE or the Authorized local technical service.
WARNING!! Risk of injuries for the patient
When using additional components in the respiratory systems or
configurations not conform to those supplied with the lung ventilator, the
inspiratory and expiratory resistance can increase until overcome the
standard requirements.
When using configurations of this type, it is necessary to pay particular
warning to the measuring values.
WARNING!! Risk of electric shock.
In case of malfunctioning of ground conductor, the connection of other
electric equipment to the supplementary outlet of the lung ventilator could
cause an increase in leakage current over the values allowed by the law.
In case of connection of other devices to the supplementary outlet, it is
necessary to verify the total leakage current.
In case of overcoming of the values allowed for total leakage current, do
not connect other devices to the supplementary outlet of lung ventilator;
connect other devices to a separate electric outlet.
All the system must satisfy the requirements for electromedical equipment
stated by the IEC/EN 60601-1-1 and IEC/EN 60601-1-2. directives.