Mount the cover of valves group.
•
Close the left front panel of the trolley.
3.2.2 Battery charger
•
In case of electric power failure, the lung ventilator is equipped with a internal
batteries that guarantees (if perfectly efficient) at least 90 minutes operation (with
default physiological respiratory parameters).
•
The switching to battery operation is made automatically: on the lung ventilator screen
appears the relevant message " POWER SUPPLY FAULT ".
•
The lung ventilator battery can be recharged by connecting the equipment to main
power supply (using power cable supplied with the unit) and placing the main switch in
" I " position.
3-6
WARNING!! Risk of injury for the patient
Check of electric connection of O 2 sensor.
At lung ventilator start-up, the system check the presence of the electric
connection to the O 2 cell ( "SELF TEST" phase ).
Carry a battery charge at least 8 hours, before using the lung ventilator the
first time.
The operating time of lung ventilator guaranteed by the battery, can vary in
the following cases:
•
old battery or not perfectly efficient;
•
not standard ventilatory parameters.
User manual, DU30651043