4.12.1 SETUP options in MAIN MENU

English, French, Spanish, Italian,
German, etcc....
Line: The graphics line is traced
without filling
Area: the graphics line is filled
Power saving mode that activates
after 30 minutes, if no command is
selected and the keyboard and
encoder knob are not used
ADULT, PAEDIATRIC, NEONATAL
Siaretron4000e
The PATIENT TYPE choice is showed on the display of the lung ventilator
together with the selected Operative Mode.
Selection of the lung ventilator display
language.
Selection of the type of graphic displayed
during respiratory phase: Square or Line.
Regulation of the alarm signal acoustic
intensity: 1 - 20
Regulation of the machine power consumption
percentage: 0 - 100.
If set to 0, when activating this function, the
display will show a black screen.
Regulation of the display brightness: 1 - 30
Setting of the patient type connected to the lung
ventilator.
The PATIENT TYPE selection activates the
default physiological breathing parameters
(PRF) and alarms (pre-set at factory) according
to the Operative Mode selected ( see on 4.3 ).