6.4.2 Disposable bacteria filter

Disposable
bacteria filter
6-10
CAUTION
It is important to use a disposable bacteria filter at the expiratory port of
ventilator between the patient circuit's expiratory limb and the expiratory
flow sensor.
Do not clean or re-use if the filters are the
disposable type.
WARNING !!
Risk of device failure
Disposable bacteria filter.
Do not attempt to sterilize and reuse a disposable bacterial /HEPA filter.
It is meant for single use.
Do not use filters that aren't in sealed pack.
Do not try to dry the filter and reuse in the same patient in case of
increased resistance.
Do not use it longer than the time-period recommended by
manufacturer.
Do not use filters that had exceeded the expiry date.
Disposable bacteria filter.
Replace the bacterial filter as per manufacturer's advice (24 - 72 hours).
If used with humidifier, change the filter more frequently (as and when
the resistance increases or patient's "work of breathing" increases).
Immediately replace the filter if it is stained with blood, secretions or
other human fluid
Check the expiry date before using on patient.
Change the filter early if nebulisation is done frequently.
Components that cannot be
destroyed should be sterilized
and disinfected according to
local standards.