O2 Sensor Calibration; Leak Test - Siare Siaretron 4000 User Manual

Icu ventilator
Hide thumbs Also See for Siaretron 4000:
Table of Contents

Advertisement

3.9.2 O
sensor calibration
2

3.9.3 Leak Test

3-52
TEST ABORTED
If the system fails the test, please see chapter Alarms - Trouble Shooting or
contact the nearest Siare Support Centre or any other support centres
authorised by Siare.
At the end of the respiratory flow sensors calibration the system
come back to the MAIN MENU.
In case of first calibration of ventilator, we suggest to execute this
procedure after having checked the PEEP calibration.
For more information please refer to the SERVICE manual.
CAUTION
For additional informations on the " TEST ON DEMAND " not mentioned in
this chapter, please refer to 3.6.2.
CAUTION
For additional informations on the " TEST ON DEMAND " not mentioned in
this chapter, please refer to 3.6.2.
User manual, DU30651043