Preliminary Checks - Alarms - Siare Siaretron 4000 User Manual

Icu ventilator
Hide thumbs Also See for Siaretron 4000:
Table of Contents

Advertisement

3.7.8 Preliminary
3-42
Ventilator operation check
Make sure that the airways pressure increases during the inspiratory
phase.
Make sure that the airways limit pressure intervenes (pressometric
operating mode).
Make sure that the variation in the set oxygen concentration value (O 2
%) corresponds.
Make sure that the lung ventilator responds properly at parameters
variation.
Make sure that that the trigger works properly.
Make sure that the values set for respiratory frequency and volume are
properly displayed and the pressure, volume and flow curves match the
monitored parameters.
Make sure that the alarms intervene properly.
If the O 2 measured value differs from the set value by more than +/-
10%, please repeat the "O 2 Sensor Calibration" procedure: TEST ON
DEMAND.
If the Vte measured value differs from the set value by more than +/-
20% ( ADULTS parameters ), please go to the "CALIBRATION
PROGRAMS" visualization to perform the Calibration of the expiratory
flow sensor (see chapter 3.8 and 4.13).
checks - Alarms
To carry out the preliminary checks relative to lung ventilator's alarms, you
need to know how the keyboard and the alarms work: please refer to the
relative chapters.
User manual, DU30651043