3.7.7 Preliminary checks - Parameters monitoring

Siaretron4000e
Based on the PRP set by the clinician and on the patient's [patient
simulator] characteristics, the lung ventilator is able to monitor and
measure a series of values necessary for the patient's clinical
evaluation.
Before checking the value of the set parameters, leave the lung
ventilator on for at least 15 minutes. This way the system will be able to
reach its operating condition.
5) Check the compliance
between the parameters set
and those monitored.
In the middle of the screen the
system displays the operating
curves.
On the top of the screen you
can see the monitored
parameters.
6) Change the PRP values.
PEEP : 5, 10 cmH 2 O
Tr. I : -1 cmH 2 O, 1 L/min
O 2 : 60%
7) Check the correspondence
between the monitored
parameters and the displayed
curves.