6.5.2 Cleaning, disinfection and sterilization before use with another patient

We recommend the use of procedures for sterilization and disinfection referred to in the
preceding paragraphs when a new patient must use the lung ventilator.
6.6

Repairs and spare parts

6.6.1 Spare parts kit for lung ventilator

Siaretron4000e
WARNING !!
Risk of injury for the patient ( * )
All maintenance and/or repair operations require perfect knowledge of the
ventilator and must therefore only be carried out by highly qualified
personnel, specifically trained and formally authorised by SIARE.
Inappropriate intervention or unauthorised modifications can compromise
safety and cause danger to the patient.
CAUTION
To avoid damage to components due to excessive wear, carry out
preventive maintenance and replace parts following the recommended
frequency.
WARNING !!
Risk of injury for the patient
It is recommended to sterilize / disinfect the lung ventilator every time is
used with another patient.
Use only original SIARE spare parts or spare parts checked and approved
by SIARE.
Code : R041000A1
Spare parts kit for annual maintenance to be used with the Siaretron 4000
lung ventilator for Adult patients: code 960401
Code : R043000A1
Spare parts kit for annual maintenance to be used with the Siaretron 4000
lung ventilator for Paediatric and Neonatal patients, code 960403.
6-13