2.4.3 Operative mode

Siaretron4000e
You can select one of these ventilation modes using the SET function
(please see 2.4.2 or 2.6).
Assisted pressure controlled ventilation, synchronised with
patient's breathing with leak compensation.
(Volume Targeted) Assisted pressure controlled ventilation,
synchronised with patient's breathing and with assured current
volume.
Assisted pressure support ventilation with assured respiratory
rate set by the User (Apnea Back Up) with leak compensation.
(Volume Targeted) Pressure support ventilation with assured
current volume and assured safety respiratory rate set by the
User (Apnea Back-Up).
Volume targeted controlled ventilation synchronised with the
patient if the inspiratory trigger is activated.
Volume targeted controlled ventilation synchronised with the
patient if the inspiratory trigger for paediatric and neonatal
patients is activated.
The VC-VAC-BABY Operative Mode is not active when the PATIENT TYPE
selected is: ADULT
Volume-targeted synchronised intermittent mandatory
ventilation.
2-13