Annex C - Table 3 - Siare Siaretron 4000 User Manual

Icu ventilator
Hide thumbs Also See for Siaretron 4000:
Table of Contents

Advertisement

A.5.3 Annex C - Table 3
Guidance and manufacturer's declaration – electromagnetic immunity
The Siaretron 4000 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Siaretron 4000 should assure that it is used in such an environment.
IMMUNITY test
3 Vrms 150 kHz to
80 MHz outside
ISM bands
Conducted RF
IEC 61000-4-6
10 Vrms 150 kHz
to 80 MHz in ISM
bands
Radiated RF
10 V/m 80 MHz to
IEC 61000-4-3
2,5 GHz
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Note:
1. At 80 MHz and 800 MHz, the higher frequency range applies.
2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
A-20
IEC 60601 TEST
Compliance
LEVEL
level effective
VRMS
VRMS
Compliance
level
[V1] VRMS
[V2] VRMS
V/m
[E1] V/m
Electromagnetic environment –
guidance
Recommended separation
distances
SEE ANNEX E
SEE ANNEX E
80 ÷ 800 MHz
SEE ANNEX E
800 ÷ 2500 MHz
SEE ANNEX E
User manual, DU3065104