3.7.2 O 2 Sensor calibration
TESTS ON DEMAND. O
function is selected.
3-32
CAUTION - TESTS ON DEMAND
The TESTS ON DEMAND screen shows various tests.
Respiratory Flow Sensors Calibration
O
Sensor Calibration
2
Leak Test
Gas Sensor : Zero Reference Calibration (optional device)
On this " TEST ON DEMAND " phase you should consider the O
calibration and Leak Test only; while for the other tests please refer to
chapter 3.9.
WARNING !! Ventilator malfunctions risk
This procedure should be carried out to check the proper operation
of the O
sensor.
2
To avoid any patient injury risks, carry out this procedure monthly.
sensor calibration
2
Press the knob selecting,
TESTS ON DEMAND.
The TESTS ON DEMAND
screen appears.
User manual, DU30651043
Sensor
2