5.3.4 Alarms DEFAULT values

Alarm Settings
Pressure (cmH 2 O)
Respiratory Rate
(bpm)
Tidal Volume (ml)
Minute Volume (L)
PEEP (cmH 2 O)
FiO 2 (%)
Gas sensor
Power Failure
Apnea Time
(configurable in MAIN
MENU – SETUP area)
5-14
Wait for a few seconds, the system automatically returns to STAND-BY
display.
Press ON-OFF key to return to STAND-BY visualization.
Adult
5 - 40
8 - 20
100 - 1000
2 - 12
Paediatric
5 - 35
12 - 30
50 - 30
1 - 8
0 - 15
21 - 80
1.0 - 5.0
Enable
20 sec
User manual, DU30651043
Neonatal
5 - 35
25 - 50
10 - 30
1 - 4