6. When water is visible from underneath the lid's top plate, observe that
water is flowing freely from the 6 different adapters.
7. Repress the "Down"-Key.
8. Remove the black extension piece for water from the chamber and
the lid.
9. Repeat the process for all available Standard and Flex lids.
IMPORTANT
If the water does not flow as described above, please contact your
authorized dealer.
6.4

Checking oil supply to the Standard lid

Frequency: weekly or every 50 cycles (whichever comes first).
In order to make sure that the DAC Universal is able to lubricate the
instruments correctly, the flow through the internal oil channels of the
Standard lid needs to be tested.
IMPORTANT
The oil supply to the lid needs only to be checked for the Standard lid,
since this is the only one with internal oil channels.
IMPORTANT
This test will use water to test the internal oil channels, not oil.
The DAC Universal is switched on and ready for operation.
The lid of the DAC Universal is removed and there is an unobstructed
view into the chamber.
You are wearing your protective gear.
The Standard lid and the white lid extension piece for oil are at hand.
The Standard lid and the chamber are cooled down.
1. Attach the white lid extension piece for oil to the bottom of the lid.
2. Place the other end of the white lid extension piece for oil on the
bottom adapter in the bottom of the chamber.
3. Assure the lid is attached vertically on the bottom adapter, as shown.
6 Cleaning and maintenance
6.4 Checking oil supply to the Standard lid
107