6 Cleaning and maintenance
6.8 Removing and cleaning the drain tank
114
2. Remove the condensation tube.
3. Empty the drain tank. Dispose of the waste water according to the
national guidelines and the applicable regulations for waste disposal.
4. Rinse the drain tank thoroughly with hot water and then empty it again
completely.
5. Soak a soft cloth in hot water and wring it out well. Use the cloth to
clean the exterior of the drain tank and the condensation tube.
6. Fill the clean drain tank with water up to the lowest mark on the label.
Sirona Dental A/S