Additional Indications; Creating A Further 445Nm And A 660 Nm Program; Examples Of Treatment Risk - Dentsply Sirona SiroLaser Blue Operating Instructions Manual

Hide thumbs Also See for SiroLaser Blue:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions SiroLaser Blue
Application
Used
wave-
length
[nm]
Treatment of
970
canker sores
66 56 842 D3648
D3648.201.03.02.02    10.2018

Additional indications

6.3
Power
Modus Time
[W]
[sec]
2.0
CW
cont.

Creating a further 445nm and a 660 nm program

6.4
The device has one general preset program for 445nm and no preset
programs for 660 nm. To create a 445nm or a 660 nm program please
create a new program in "my applications" as described in chapter My
Applications  [→ 48] and choose the 445nm or 660 nm wavelength.
6.5

Examples of treatment risk

Surgery area
Risk: Soft and hard tissue necrosis or overheating of the tooth.
Countermeasure: Use the laser beam like a scalpel, holding it
perpendicular to the surface under treatment, and never aim it at a
single point for a longer period of time. Do not select excessively high
power levels for the laser.
WARNING
Never treat perpendicular to any bone surface.
Periodontics area
Risk: Minor necrosis or scarring of the radicular area.
Countermeasure: When working in periodontal pockets, always aim the
laser parallel to, i.e. never perpendicular to, the roots. Run the distal
end of the optical fiber over the entire inner surface of the periodontal
pocket.
6 Indications, contraindications and medical precautions
Duty
Fre-
Fiber
Cycle
quenz
[µm]
[%]
y
[Hz]
0
0
320
6.3 Additional indications
Help Menu
Anesthetics not needed! Apply
laser 1-3 mm away from lesion for
a few seconds - semicontact,
wave the
laser fiber over the entire lesion.
Interrupt treatment briefly, if pain
sensations occurs.