Dentsply Sirona DAC Universal Operating Instructions Manual page 129

Hide thumbs Also See for DAC Universal:
Table of Contents

Advertisement

Sirona Dental A/S
Operating Instructions DAC Universal
Code
Pictogram
no.
60
61
62
64
65
66
66 15 574 D3472
D3472.201.20.01.02 03.2019
Description
The chamber
temperature and the
theoretical steam
temperature deviate from
each other more than 2K.
The software has not
registered a change in
temperature in the steam
generator for 3 minutes.
The temperature in the
chamber did not reach
110 °C / 230 °F.
The safety system was
activated.
The safety system is
defective.
The temperature was too
high during sterilization.
7.1 Display messages and what they mean
Corrective action
Please contact your authorized dealer.
Please contact your authorized dealer.
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
If the DAC Universal is in a room with a temperature
< 16 °C / 61 °F, ensure that operating conditions are
fulfilled.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC Universal.
If the error occurs repeatedly, please contact your
authorized dealer.
Please contact your authorized dealer.
Please contact your authorized dealer.
Please contact your authorized dealer.
7 Errors / Display messages
129