5 Operation
5.1 Regular activities
52
4. Fold the chemical indicator into a V-shape.
5. Put the chemical indicator into the PCD test device.
6. Twist the PCD test device clockwise to close.
IMPORTANT
Make sure the PCD test device is completely closed.
A proper check of the PCD test device is not ensured if it is partly open.
7. Press the PCD test device onto the ISO/ INTRAmatic adapter until
you hear and feel a slight click.
8. Put instruments onto the other 5 adapters as usual.
9. Pull lightly on the instruments to ensure that they are firmly attached
to the adapters.
10. Start a standard cycle. See "Reprocessing with the Standard lid
[ → 90]" for details of the procedure.
CAUTION
Risk of burns!
Allow the extremely hot instruments to cool down after the end of the
cycle.
Sirona Dental A/S
Operating Instructions DAC Universal
66 15 574 D3472
D3472.201.20.01.02 03.2019