Sirona Dental A/S
Operating Instructions DAC Universal
66 15 574 D3472
D3472.201.20.01.02 03.2019
5. Insert the Standard lid in the holder of the DAC Universal in such a
way that the instruments are pointing down.
NOTICE
Make sure that the lid is inserted all the way into the holder so that the
instrument heads do not touch the wall of the chamber. Make sure there
are no objects in the chamber.
NOTICE
The DAC Universal and/or the instruments can be damaged if a lid is not
inserted correctly.
 There are instruments on the Standard lid; the DAC Universal is
still open and ready for a cycle.
 The display shows the last used program:
"Ready 134 °C 18min"
IMPORTANT
If you do not select a program, the last used program will be used.
If you want to change the program, proceed as follows:
The Standard lid is inserted in the holder of the DAC Universal,
➢ Enter the "Program selection" menu and use the "Left/Right"-Keys
accordingly to select the appropriate program.
➢ Press the "Down"-Key to confirm your selection.
➢ Press the "Up"-Key to return to the start screen.
 The selected program is displayed at the top of the ready screen.
The program is shown as temperature/holding time.
6. Make sure the Standard lid is selected. If not, press the "Up/Down"-
Keys accordingly, to select this lid.
7. Press the "Start/Stop"-Key to start the cycle.
5.3 Reprocessing with the DAC Universal
"Ready 134 °C"
.
5 Operation