5 Operation
5.2 Using the menu and display information
64
Check & Clean needed - is displayed simultaneously with a 2 second
acoustic alarm. Appears after every 50 runs.
Indicates that the cleaning and maintenance in accordance with the
Check & Clean guide must be performed.
Press the "Down"-Key to confirm.
For cleaning and maintenance instructions, see „Cleaning and
maintenance [ → 104]".
User maintenance needed - displays every 200 runs, simultaneously with
a 2 second acoustic alarm.
Indicates that User Maintenance needs to be performed. For cleaning
and maintenance instructions, see "Cleaning and maintenance [ → 104]".
Danger: The chamber, lid and instruments are hot. Please be careful
when opening the chamber completely by pressing the "Down"-Key.