6 Cleaning and maintenance
6.10 Cleaning the empty chamber
116
2. Press the "Down"-Key to turn on automatic water.
 "Automatic water" is now turned on and the ready screen
reappears.
IMPORTANT
If there is no permanent water connection, the DAC Universal will jump
directly to the ready screen when the DAC Universal is switched on
again.
6.10

Cleaning the empty chamber

Frequency: weekly or every 50 cycles (whichever comes first).
IMPORTANT
If there is no permanent water connection, fill the water tank with
demineralized or distilled water (water conductivity < 3.0 µS/cm).
The DAC Universal is switched on and ready for operation.
The Standard lid of the DAC Universal is removed and there is an
unobstructed view into the chamber.
There is no water or oil in the chamber.
1. Clean the empty chamber with a lint-free cloth and a solution of 70 %
isopropyl alcohol.
IMPORTANT
The chamber must not be cleaned with agents containing chlorine.
2. Insert the Standard lid in the holder of the DAC Universal.
NOTICE
No instruments must be on the Standard lid.
NOTICE
Make sure that the lid is inserted all the way into the holder.