12. Press down on the safety ring to start the cycle. If the safety ring is
not pressed within 20 seconds, you must choose the lid and press the
"Start"-Key again.
 The Basket lid of the DAC Universal closes as soon as the safety
ring is pressed.
 The DAC Universal automatically runs the cleaning and
sterilization cycle.
 The display shows at any time which phase the DAC Universal is
in, see section "Display information [ → 59]".
 When the selected cycle is completed the display will indicate as
seen on the picture.
 The lid moves about halfway out of the chamber. This indicates
that a cycle has been completed and the instruments have been
sterilized.
 The cycle number is shown next to the completed cycle symbol.
CAUTION
Risk of burns!
Never hold on to the lid or hold your head directly over the DAC
Universal when it opens, as steam will emerge from the device.
WARNING
After completion of the cycle, make sure that all solid instruments are
still in the basket. If instruments have fallen into the chamber during the
cycle, cleaning and sterilization have failed. In this case, reprocessing of
the instruments must be repeated.
13. Press the "Down"-Key.
 The Basket lid will open completely.
CAUTION
Risk of burns!
Even when the Basket lid is half or completely open, the instruments are
still hot. Allow the extremely hot instruments to cool down after the
process.
CAUTION
Risk of burns!
The metal part of the lid is very hot. At the end of the cycle, touch only
the plastic part of the lid.
14. Remove the Basket lid from the holder on the DAC Universal at the
end of the cycle to allow the instruments to cool down faster.