5 Operation
5.2 Using the menu and display information
88
5.2.5

Deactivating lubrication

It is possible to deactivate lubrication for a single cycle. This is necessary
when reprocessing a load with the Standard lid which solely contains
instruments that do not require lubrication. For all other lid types,
lubrication is automatically deactivated.
For more information on how to deactivate lubrication manually, see
"Start cycle on the Standard lid [ → 90]".
When the cycle for which lubrication was deactivated has been
completed, the lubrication is automatically reactivated.
5.2.6

Drain the chamber

If there is too much water left in the chamber, for instance from performing
Check & Clean, you can drain the chamber via the menu:
The DAC Universal is switched on and ready for operation.
The empty Standard lid is at hand and cooled down.
1. Insert the Standard lid in the holder of the DAC Universal.
2. Press the "Right"-Key to enter the right hand menu.
3. Press the "Right"-Key until the "Manual control" menu is selected.
4. Press the "Down"-Key to enter the "Manual control" menu.
5. Press the "Right"-Key until "Drain chamber" is selected.
6. Press the "Down"-Key to confirm.
 The lid closes automatically.
 The water is pumped from the chamber into the drain tank / drainage
system.
 The lid opens again automatically after 60 seconds.
IMPORTANT
If the chamber is not completely empty, repeat the process until the
chamber is completely empty.
Sirona Dental A/S
Operating Instructions DAC Universal
66 15 574 D3472
D3472.201.20.01.02 03.2019