Permanent Water Connection; Manual Filling Of The Water Tank; Drain Connection - Dentsply Sirona DAC Universal Operating Instructions Manual

Hide thumbs Also See for DAC Universal:
Table of Contents

Advertisement

4 Installation
4.2 Installation of connections
42
● Permanent water connection.
The DAC Universal is connected to a permanent water conditioning
system. NitraDem Direct Connect is recommended as water supply
for the DAC Universal.
● Manual filling of the water tank with demineralized / distilled water,
see "Filling the water tank [ → 56]".

Permanent water connection

4.2.3.1
IMPORTANT
Observe local requlations for backflow protection when connecting the
DAC Universal to permanent water supply. According to EN1717 an air
gap is required as backflow protection of what EN1717 specifies as
being Category 5 fluids.
The following must be observed when connecting the DAC Universal to a
permanent water supply:
To operate the DAC Universal with a water conditioning unit, it must be
switched to "Automatic water supply". This is done via the menu system,
see "Menu system. [ → 65]".
If you use a permanent water supply and have not installed a siphon,
please empty the drain tank once a day or whenever it is full, see
"Removing and cleaning the drain tank [ → 113]".
NOTICE
A water filter (20 µm) must be installed upstream of the DAC Universal
to ensure that the water is clean.

Manual filling of the water tank

4.2.3.2
If no permanent water connection is selected, the integrated water tank
must be filled with water manually.
See "Filling the water tank [ → 56]" for the procedure.
4.2.4

Drain connection

NOTICE
Do not start a cycle if the DAC Universal is not connected to the drain
tank or drainage system.
The DAC Universal works only in a one-way water system, i.e. used water
(waste water) is not reused.
Therefore, the "Drain" connection on the back of the DAC Universal must
be connected either to:
● a drain tank, see "Connecting the drain tank [ → 43]" or
● a siphon, see "Connecting the siphon [ → 44]"
Sirona Dental A/S
Operating Instructions DAC Universal
66 15 574 D3472
D3472.201.20.01.02 03.2019