7 Errors / Display messages
7.1 Display messages and what they mean
Code
Pictogram
no.
23
24
25
26
27
124
7.1

Display messages and what they mean

The display messages each have an assigned number. The number will
be visible in the top left corner, next to the error symbol (Not necessarily
shown below). The unique code number for each display message is
shown in the table below, combined with an explanation to the specific
display message:
Description
The heat capacity of the
steam generator was not
regained during the cycle.
Venting error.
Venting error.
Venting error.
Venting error.
Operating Instructions DAC Universal
Corrective action
If the DAC Universal is in a room with a temperature
< 16 °C / 61 °F, ensure that operating conditions are
fulfilled.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC Universal.
If the error occurs repeatedly, please contact your
authorized dealer.
Please contact your authorized dealer.
Please contact your authorized dealer.
Please contact your authorized dealer.
Please contact your authorized dealer.